LifeVest 4000 Patient Manual
Device disabled message
If the LifeVest detects an untreatable condition called
asystole, where your heart slows and stops, it will display
this message to alert bystanders.
asystole, where your heart slows and stops, it will display
this message to alert bystanders.
Under rare circumstances, it may display this message
when your heart has not stopped.
when your heart has not stopped.
If you are conscious and you feel OK and you get this
message, follow the instructions below.
message, follow the instructions below.
If you get this message and you’re conscious
The problem may be with the signal from the ECG electrodes. Follow this
procedure to correct the problem.
procedure to correct the problem.
1 Remove and reinstall the battery.
2 If the message returns, check your electrode belt and garment. Make sure:
y
All ECG electrodes are touching your skin.
y
Nothing is between the ECG electrodes and your skin, such as clothing or
one of the cables.
y
Garment and belt fit snugly, with the electrodes pressing against your
skin.
3 If you continue to get this message even though you remain conscious, call
ZOLL Lifecor.
Page 5-18
1-800-543-3267