Instructions for Use | Therapy Modes and Controls 
 
 
 
Therapy Control Settings 
28 
 
Setting Name 
Setting Range/Increment 
• 
All circuits: 0-20 cmH
2
O  
• 
Increments: 1 cmH
2
Breath Rate 
Adult patient type: 0-80 BPM, 1BPM increments 
Pediatric patient type: 0-60 BPM, 1BPM increments 
Infant patient type: 0-40 BPM, 1 BPM increments 
CPAP 
Adult and pediatric patient types: 3-25 cmH
2
O, 1 cm H
2
O increments 
Infant patient type: 3-15 cmH
2
O, 1 cm H
2
O increments 
EPAP Min/Max 
Adult and pediatric patient types: 3-25 cmH
2
O, 1 cm H
2
O increments 
Infant patient type: 3-15 cmH
2
O, 1 cm H
2
O increments 
FiO
2
 
21-100%, 1% increments 
(21% = ambient condition, no control) 
Flow Cycle Sensitivity  10-90%, 1% increments 
Flow Pattern 
Square: airflow is constant 
Ramp: inspiration airflow starts high and decreases 
Flow Trigger Sensitivity  0.5 (high sensitivity) to 9 L/min (low sensitivity) 
Inspiratory Time 
Adult patient type: 0.5-5.0 seconds, 0.1 second increments 
Pediatric patient type: 0.3-2.0 seconds, 0.1 second increments 
Infant patient type: 0.3-1.0 seconds, 0.1 second increments 
IPAP  
Adult patient type: 3-60 cmH
2
O, 1 cmH
2
O increments 
Pediatric patient type: 3-45 cmH
2
O, 1 cmH
2
O increments 
Infant patient type: 3-35 cmH
2
O, 1 cmH
2
O increments 
Max Pressure 
• 
6-60 cmH
2
O, 1 cmH
2
O increments 
PEEP 
Adult patient type:  
• 
Active circuit: 0-35 cmH
2
• 
Passive circuit: 3-25 cmH
2
• 
Increments: 1 cmH
2
Pediatric patient type: 
• 
Active circuit: 0-25 cmH
2
• 
Passive circuit: 3-25 cmH
2
• 
Increments: 1 cmH
2
Infant patient type: 
• 
Active circuit: 0-15 cmH
2
• 
Passive circuit: 3-15 cmH
2
• 
Increments: 1 cmH
2
Pressure Control 
Adult patient type:  
• 
All circuits but passive: 0-60 cmH
2
O  
• 
Passive circuit: 0-57 cmH
2
• 
Increments: 1 cmH
2
Pediatric patient type:  
• 
All circuits: 0-30 cmH
2
O  
• 
Increments: 1 cmH
2
Infant patient type:  
• 
All circuits: 0-20 cmH
2
O  
• 
Increments: 1 cmH
2
Pressure Support 
Adult patient type:  
• 
All circuits but passive: 0-60 cmH
2
• 
Passive circuit: 0-57 cmH
2
• 
Increments: 1 cmH
2
Pediatric patient type:  
• 
All circuits: 0-30 cmH
2
O  
• 
Increments: 1 cmH
2
Infant patient type:  
• 
All circuits: 0-20 cmH
2
O  
• 
Increments: 1 cmH
2
PS Min/Max 
• 
All circuits but passive: 0-60 cmH
2
• 
Passive circuit: 0-57 cmH
2
• 
Increments: 1 cmH
2
Rise Time 
0 (faster) to 6 (slower) 
Tidal Volume 
Adult patient type: 
70-1200 ml 
Increment: 5 ml