Instructions for Use | Therapy Modes and Controls 
 
 
 
Therapy Control Settings 
27 
 
Therapy Actions 
Oxygen Flush 
D
ESCRIPTION
 
This feature requires the oxygen blender. When active, the device delivers 100% oxygen for two minutes. This feature functions independent of 
any oxygen blending setting. During an oxygen flush, the High FiO
2
 alarm is disabled. 
W
ORKING WITH 
O
XYGEN 
F
LUSH
 
To start an oxygen flush:  
Tap 100% O
2
 in the status bar and then tap Start. A timer appears that counts down 
the two minutes.  
 
To stop an oxygen flush:  
Tap 100% O
2
 in the status bar and then tap Stop. 
 
 
 
A
PPLICABLE 
T
HERAPY 
M
ODES
 
All 
 
Manual Breath 
D
ESCRIPTION
 
Delivers a breath based on the current therapy mode settings.  
To deliver a manual breath:  
Tap the Manual Breath button in the status bar and then tap Start.  
 
 
A
PPLICABLE 
T
HERAPY 
M
ODES
 
• 
A/C-PC 
• 
A/C-VC 
• 
PSV 
• 
S/T 
• 
SIMV-PC 
• 
SIMV-VC 
 
 
Therapy Control Settings 
Therapy control settings can be interdependent. For guidance, see the previous therapy mode descriptions. 
Setting Name 
Setting Range/Increment 
Backup Pressure 
Support 
Adult patient type:  
• 
All circuits but passive: 0-60 cmH
2
• 
Passive circuit: 0-57 cmH
2
• 
Increments: 1 cmH
2
Pediatric patient type:  
• 
All circuits: 0-30 cmH
2
O  
• 
Increments: 1 cmH
2
Infant patient type: