Respironics Inc. 1127941 Manual Do Utilizador
Instructions for Use | Therapy Modes and Controls
Therapy Control Settings
27
Therapy Actions
Oxygen Flush
D
ESCRIPTION
This feature requires the oxygen blender. When active, the device delivers 100% oxygen for two minutes. This feature functions independent of
any oxygen blending setting. During an oxygen flush, the High FiO
any oxygen blending setting. During an oxygen flush, the High FiO
2
alarm is disabled.
W
ORKING WITH
O
XYGEN
F
LUSH
To start an oxygen flush:
Tap 100% O
Tap 100% O
2
in the status bar and then tap Start. A timer appears that counts down
the two minutes.
To stop an oxygen flush:
Tap 100% O
To stop an oxygen flush:
Tap 100% O
2
in the status bar and then tap Stop.
A
PPLICABLE
T
HERAPY
M
ODES
All
Manual Breath
D
ESCRIPTION
Delivers a breath based on the current therapy mode settings.
To deliver a manual breath:
Tap the Manual Breath button in the status bar and then tap Start.
Tap the Manual Breath button in the status bar and then tap Start.
A
PPLICABLE
T
HERAPY
M
ODES
•
A/C-PC
•
A/C-VC
•
PSV
•
S/T
•
SIMV-PC
•
SIMV-VC
Therapy Control Settings
Therapy control settings can be interdependent. For guidance, see the previous therapy mode descriptions.
Setting Name
Setting Range/Increment
Backup Pressure
Support
Adult patient type:
•
All circuits but passive: 0-60 cmH
2
O
•
Passive circuit: 0-57 cmH
2
O
•
Increments: 1 cmH
2
O
Pediatric patient type:
•
All circuits: 0-30 cmH
2
O
•
Increments: 1 cmH
2
O
Infant patient type: