Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Device Setup 
 
 
 
Starting Trilogy Evo Universal 
36 
 
Low Flow Oxygen 
To connect low flow oxygen:  
1.  Connect the oxygen tubing to the O
2
 adapter supplied with the device.  
2.  Connect the O
2
 adapter to the low flow oxygen inlet on the Utility Panel by pressing down on the valve.  
 
 
 
 
Oxygen-Related Alarms  
After the oxygen is connected, set the oxygen-related alarms. See chapter 6, “Alarms and System Messages” for instructions.  
Oxygen alarms are as follows: 
• 
Oxygen Regulation 
• 
Low SpO
2
 
• 
High SpO
2
 
• 
Low FiO
2
 
• 
High FiO
2
 
• 
Low Oxygen Input Pressure 
• 
High Oxygen Input Pressure 
 
Starting Trilogy Evo Universal  
To start Trilogy Evo Universal: 
1.  Visually inspect Trilogy Evo Universal and all accessories, cords, and tubes attached to the device. 
2.  Verify that circuit connections are secure. 
3.  Press the On/Off (Standby) button.  
4.  Listen for a minimum of three beeps as Trilogy Evo Universal performs system start up checks. The beeps test all alarm signals to ensure 
proper functioning.  Ensure no system messages appear.  
5.  Watch as the light bar and Alarm Silence button blink once red and once yellow.  
6.  Confirm that the power sources you have connected are functioning properly and that power is sufficient. For help, see the “Power 
Management” chapter.