Instructions for Use | Device Operation 
 
 
 
About Prescriptions 
37 
 
5. 
Device Operation
 
 
Clinical Assessment 
Warning 
Before placing a patient on the ventilator, perform a clinical assessment. Considerations should include:
 
• 
Choosing alarm settings 
• 
Assessing whether alternative ventilation equipment is required 
• 
Selecting additional accessories, including the patient monitoring accessories you will use 
For ventilator-dependent patients, always have alternate ventilation equipment, such as a back-up ventilator or manual resuscitator available.  
Ventilator-dependent patients should be continuously monitored by qualified personnel. These personnel should be prepared to provide alternate 
therapy in the event of ventilator failure or inoperative equipment. 
Entering New Patient Information 
To enter new patient information 
1.  In the Home window, tap the New Patient button. This button clears all existing patient data. 
2.  In the New Patient window, select a Patient Type
• 
Infant 
• 
Pediatric 
• 
Adult 
3.  Select the Patient Sex
4.  For infant patients, in the Weight section, use the slider or the plus and minus buttons to select the patient’s weight.  
For pediatric or adult patients, in the Height section, select the patient’s height. 
Note: this information is used to establish default therapy and alarm settings, including tidal volume and alarms based on tidal volume. 
5.  In the title bar, tap Accept to save your choices.  
6.  Acknowledge the reminder to ensure a viral/bacterial filter is installed on the outlet of the device. 
7.  Edit the prescription settings according to the procedure in the next section, “About Prescriptions.” Therapy and alarm settings differ 
based on Patient Type. See the “Therapy Modes and Controls” and “Alarms” chapters. 
About Prescriptions 
System Timeout 
When working in the Prescription Settings window, ensure you save your changes. After a period of inactivity, the system reverts to the previous 
setting and your changes are not saved. If you are making changes during therapy delivery, the period of inactivity is 30 seconds. If you are making 
changes when the system is in standby, the period of inactivity is 5 minutes.