Instructions for Use | Alarms and System Messages 
 
 
 
About Alarms 
41 
 
6.  Alarms and System Messages 
Overview 
Trilogy Evo Universal generates audible and visual alarms to alert you when conditions require attention. 
Warning: To prevent death or serious injury, monitor the patient and the ventilator regularly in order to determine the need to provide emergency 
ventilation when an alarm sounds or the ventilator malfunctions. Always test the alarms after changing the circuit or prescription. 
An increase in circuit resistance can prevent proper operation of some alarms. 
Speaking valves, heat moisture exchangers (HMEs), and filters create additional circuit resistance and may affect the performance of 
alarms chosen for circuit disconnect protection. 
Alarm settings are retained when power is lost. 
Do not rely on any single alarm to detect a disconnected circuit.  Certain components may affect the performance of the alarms chosen 
to signal that a circuit is disconnected.  Use the apnea, low tidal volume, low minute ventilation, and low respiratory rate alarms in 
conjunction with the circuit disconnect alarm.  Test these alarms daily and after you change ventilator settings. 
About Alarms 
When an alarm is active, the following indicators occur:   
• 
An Alarm List appears in the menu bar. 
• 
An Alarm light bar flashes red or yellow, or glows steady yellow, depending on the alarm level (To configure this indicator, see chapter 7, 
“Device Options.” 
• 
The Alarm Silence button on the device flashes red or yellow, or glows steady yellow, depending on the alarm level 
• 
An audible alarm sounds. 
 
Trilogy Evo Universal uses three alarm levels: 
• 
High Priority – Requires an immediate response 
• 
Medium Priority – Requires a prompt response 
• 
Low Priority – Requires awareness. 
Trilogy Evo Universal also shows system messages to inform you about changing conditions. These messages are described in “Alarms and System 
Messages. 
When an alarm or system message becomes active, Trilogy Evo Universal stops the screen saver and disables the automatic screen lock. 
When external monitors are used, such as an FiO
2
 sensor or SpO
2
 sensor, associated alarm settings only appear when the ventilator detects sensor 
connection. Alarm settings are stored in the system so if a sensor becomes disconnected, the alarm settings are restored upon reconnection. To 
delete external monitor alarm settings, clear all patient data according to the procedure described in the “Device Operation” chapter. 
Some patient-related alarms have configurable settings, while others have fixed settings. All system and power alarm settings are fixed. Trilogy Evo 
Universal also shows system messages to inform you about changing conditions. Alarm and system message details appear in the “Alarms and 
System Messages” section in this chapter. 
 
Alarm and Message Indictors 
Icon 
Description 
Light and sound indicators 
 
High alarm 
Light bar flashes red 
Audible alarm repeats rapidly 
 
Medium alarm 
Light bar flashes yellow 
Audible alarm repeats moderately