Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Alarms and System Messages 
 
 
 
Resetting Alarms 
43 
 
Setting the Alarm Volume 
Warning: 
Ensure the alarm volume is set loud enough for the caregiver to hear it. Consider the use of a remote alarm. If you do use a remote alarm, test it 
before starting ventilation. 
To set the alarm volume: 
1.  In the menu bar, tap the Options icon. 
  
2.  In the Options window, tap Device Settings
3.  In the Device Settings window, tap Alarm Volume.  
4.  In the setting dialog box, make your selection. 
5.  On the Alarm Volume dialog box, select the volume you want and then tap the Accept checkmark.  
Responding to an Alarm 
Warning 
Visually monitor the patient and ventilator at all times during an Alarm Silence period. Allowing alarms to continue without intervention may result 
in harm to the patient. 
When an alarm occurs: 
1.  Ensure the patient has adequate ventilation and oxygen. If required, provide an alternate method of ventilation. 
2.  Tap the Alarm List to view all alarms and messages. If you see the help icon   , you can tap it for more information.  
3.  Press the Alarm Silence button on the device to pause all audible alarms for 2 minutes.  
 
4.  Take action to resolve the alarm. For help on specific alarms, see the “Alarms and System Messages” section. 
Resetting Alarms 
 
The Alarm Reset icon resets all active and resolved alarms.
 
 
When an alarm-triggering condition is no longer present, the alarm status changes to Resolved and the alarm or message remains in the list. Tap 
the Alarm Reset icon to reset the alarm list.