Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
50 
 
Algorithm 
summary  
The flow from the device at the end of the expiratory phase is greater than a threshold based on a 
typical flow that would result if a 0.25” diameter orifice leak were present in the circuit. 
 
R
EBREATHING 
D
ETECTED
 
Priority 
Medium 
Why it occurs 
The ventilator detects the potential for the inhalation of exhaled gases.  Rebreathing is detected at 3 
consecutive breaths for passive circuits and 6 consecutive breaths for Active PAP, Active Flow, or Dual 
Limb circuits. 
What to do 
Check for a partially occluded exhalation port or increase expiratory leak flow. 
Ensure the exhalation valve is attached. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when exhaled air returns to a non-hazardous level for 6 
consecutive breaths. 
 
Device will continue to operate. 
Algorithm 
summary  
This alarm is based on an estimation of the inhaled fraction of carbon dioxide (FiCO
2
). For each breath, 
the ventilator assigns a percentage of FiCO
2.
 When the sum of the percentages for the most recent 
breaths exceeds 30%, the alarm is announced.  This method of using a running sum of concentration 
for the alarm reporting results in a shorter alarm delay when there are higher concentrations of FiCO
2
.   
The concentration estimate is based on nominal volumetric capnography curves, measured flow in the 
inhalation limb, and patient tidal volume. Any three breaths without rebreathing resets the sum of 
percentages to zero. 
 
This figure shows the assumed shape of the capnograph for a given volume of exhaled gases. 
 
 
 
 
V
OLUME 
U
NDER 
D
ELIVERY
 
Priority 
Medium 
Why it occurs 
A system limit is reached and the set volume cannot be reached for 3 consecutive breaths. 
What to do 
Check the patient. Verify that the circuit and airway are clear of obstruction.  Verify that the high 
pressure alarm limit is sufficient. Review all settings including Inspiratory Time, Tidal Volume, and Flow 
Pattern. Verify that patient and ventilator are synchronous by examining the waveforms and adjusting 
the trigger settings.  
Requirements 
The mode is A/C-VC or SIMV-VC. 
Device 
performance 
during active 
alarm 
Device will continue to operate, attempting to deliver the set therapy 
All monitored parameters and alarms will continue to function. 
Algorithm 
summary  
In modes designed to regulate the tidal volume on each breath such as AC-VC or SIMV-VC, the alarm is 
generated when the inhaled tidal volume is less than or equal to 85% of the tidal volume setting for 3 
consecutive breaths.