Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
51 
 
L
OSS OF 
CO
2
 
S
IGNAL
 
Priority 
Medium 
Why it occurs 
The EtCO
2
 sensor is reporting invalid data, the sensor signal is lost, or no breaths are detected for 
more than 10 seconds while the device is delivering therapy. 
What to do 
Ensure that the EtCO
2
 sensor is properly attached to the ventilator and the patient circuit. 
Requirements 
The High EtCO
2
 alarm or Low EtCO
2
 alarm is enabled and the sensor was previously reporting valid 
data for 3 continuous seconds.. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the EtCO
2
 sensor is properly attached to the ventilator and 
the patient and the sensor reports data. 
 
Device will continue to operate. 
 
L
OSS OF 
S
P
O
2
 
S
IGNAL
 
Priority 
Medium 
Why it occurs 
The oximeter is reporting invalid data or the oximeter is disconnected for more than 10 seconds while 
the device is delivering therapy or is in standby. 
What to do 
• 
Ensure that the SpO
2
 probe is properly attached to the ventilator and the patient. 
• 
Reposition the probe on the patient, if necessary. 
Requirements 
Low SpO
2
, high SpO
2
, low pulse rate, or high pulse rate alarm is enabled and the oximeter was 
previously reporting valid data for 3 continuous seconds. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the SpO
2
 probe is properly attached to the ventilator and to 
the patient and the oximeter reports data for more than 10 seconds. 
 
Device will continue to operate. 
 
CO
2
 
S
ENSOR 
A
DAPTER 
Z
ERO 
R
EQUIRED
 
Priority 
Medium 
Why it occurs 
The CO
2
 sensor requests a zero when the device is delivering therapy. 
What to do 
Reestablish the baseline CO
2
 level. Follow the “CO
2
 Sensor Adapter Zero” procedure in the “Device 
Options” chapter.  
Requirements 
CO
2
 sensor 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
CO
2
 monitored parameters and alarms will not function 
 
C
HECK
/C
HANGE 
CO
2
 
A
IRWAY 
A
DAPTER
 
Priority 
Medium 
Why it occurs 
The CO
2
 sensor reports that a check is required when the device is delivering therapy. 
What to do 
Check the CO
2
 sensor 
Requirements 
CO
2
 airway adapter 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
CO
2
 monitored parameters and alarms will not function 
 
CO
2
 
S
ENSOR 
F
AILURE
 
Priority 
Medium 
Why it occurs 
The CO
2
 sensor reports a fault while the device is delivering therapy. 
What to do 
• 
Disconnect and then reconnect the sensor.  
• 
Replace the sensor.  
Requirements 
CO
2
 sensor 
Device 
performance 
Device will continue to operate. 
CO
2
 monitored parameters and alarms will not function