Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
54 
 
The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume exceeds 
the low tidal volume alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
10 to 2000 ml in increments of 5 ml. 
 
H
IGH 
M
INUTE 
V
ENTILATION
 
Priority 
Medium 
Why it occurs 
The patient’s minute ventilation is greater than or equal to the high minute ventilation alarm setting.  
What to do 
Check if the patient is in pain or in an anxious state. 
 
Note: A patient in pain or in an anxious state may breathe at a higher rate or at a higher tidal volume. 
Such conditions may contribute to higher minute ventilation. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the calculated minute ventilation is less than the high 
minute ventilation alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
Adult and Pediatric patient types: 0.2 to 30 LPM in minimum increments of 0.1 LPM. 
Infant patient type: 0.2 to 10 LPM in minimum increments of 0.1 LPM. 
 
H
IGH 
R
ESPIRATORY 
R
ATE
 
Priority 
Medium 
Why it occurs 
The respiratory rate is greater than the High Respiratory Rate alarm setting.  
 
When the trigger type is ‘Off’ then the spontaneous respiratory rate will not trigger the alarm.   
What to do 
• 
Check the patient for pain or anxiety 
• 
Check the ventilator for auto-triggering 
 
Note: A patient in pain or in an anxious state may breathe at a higher rate. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured respiratory rate is less than the high 
respiratory rate alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
Off, 1 to 90 BPM in increments of 1 BPM. 
 
L
OW 
I
NSPIRATORY 
P
RESSURE 
A
LARM
 
Priority 
Medium 
Why it occurs 
The measured peak inspiratory pressure is less than or equal to the low inspiratory pressure alarm 
setting. 
What to do 
• 
Check the patient for changes that may cause this alarm, such as excessive patient inspiratory 
effort 
• 
Check the ventilator for  
o
 
Kinked, pinched, or blocked tubing 
o
 
Circuit leak or disconnect 
Requirements 
Volume Modes: 
-  CV 
-  SIMV-VC 
-  AC 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
 
The alarm is automatically resolved when the measured peak inspiratory pressure is greater than the 
low inspiratory pressure alarm setting. 
Alarm Settings 
PEEP+1 to 89 cmH
2
O in increments of 1 cmH
2
O.