Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
55 
 
L
OW 
S
P
O
2
 
Priority 
Medium 
Why it occurs 
The measured SpO
2
 is less than or equal to the Low SpO
2
 alarm setting while the device is delivering 
therapy or is in standby. 
What to do 
• 
Ensure that oxygen tubing is attached to the ventilator and the oxygen source. 
• 
Ensure that the oxygen source is providing oxygen. 
• 
Increase oxygen flow or increase FiO
2
Requirements 
Requires pulse oximeter 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured SpO
2
 rises above the Low SpO
2
 alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
50 to 95% in increments of 1%. 
 
H
IGH 
S
P
O
2
 
Priority 
Medium 
Why it occurs 
The measured SpO
2
 is greater than or equal to the High SpO
2
 alarm setting. 
What to do 
• 
Ensure that the oxygen source is appropriate. 
• 
Decrease percentage of oxygen delivery. 
Requirements 
Requires pulse oximeter 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured SpO
2
 falls below the High SpO
2
 alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
90 to 100% in increments of 1%. 
 
 
L
OW 
E
T
CO
2
 
Priority 
Medium 
Why it occurs 
The measured EtCO
2
 is less than or equal to the Low EtCO
2
 alarm setting for 10 seconds while the 
device is delivering therapy. 
What to do 
Check the patient. If there is a tracheal tube, verify it is properly inserted. Verify that leak is not 
excessive. High leak will reduce the measured EtCO
2
. Review the ventilation parameters including tidal 
volume, minute ventilation and respiratory rate and verify that they are appropriate. 
Requirements 
CO
2
 sensor 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured EtCO
2
 falls below the High EtCO
2
 alarm 
setting. 
 
Device will continue to operate. 
Alarm Settings 
1 to 100 mmHg in increments of 1 mmHg. 
 
H
IGH 
E
T
CO
2
 
Priority 
Medium 
Why it occurs 
The measured EtCO
2
 is greater than or equal to the high EtCO
2
 alarm setting for 10 seconds while the 
device is delivering therapy. 
What to do 
• 
Passive circuit: check for an insufficient leak  
• 
Active and dual limb circuits: ensure the valve is operable 
Requirements 
CO
2
 sensor 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured EtCO
2
 falls below the high EtCO
2
 alarm 
setting. 
 
Device will continue to operate. 
Alarm Settings 
1 to 100 mmHg in increments of 1 mmHg.