Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
57 
 
 
 
Low-priority Patient Alarms with Variable Settings  
L
OW 
P
ULSE 
R
ATE
 
Priority 
Low 
Why it occurs 
In the standby and therapy states, the measured pulse rate is less than or equal to the low-pulse-rate 
alarm setting. 
What to do 
Check probe placement and reposition 
Requirements 
Pulse oximeter reporting correctly for the previous three seconds 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured heart rate exceeds the low pulse rate alarm 
setting. 
 
Device will continue to operate. 
Alarm Settings 
18 to 300 beats per minute in increments of 1 beat per minute. 
 
H
IGH 
P
ULSE 
R
ATE
 
Priority 
Low 
Why it occurs 
In the standby and therapy states, the measured pulse rate is greater than or equal to the high-pulse-
rate alarm setting. 
What to do 
• 
Check oxygen flow to the ventilator 
Requirements 
Pulse oximeter reporting correctly for the previous three seconds. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured heart rate falls below the high pulse rate 
alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
18 to 300 beats per minute in increments of 1 beat per minute. 
 
 
Power Alarms 
L
OW 
B
ATTERY
 
Warning
If the high-priority “Low Battery” alarm occurs, immediately connect the ventilator to an alternate power source. If no alternate 
power source is available, immediately place the patient on an alternate source of ventilation. 
Priority 
Medium priority when the last available battery is the detachable or internal battery and can provide 
at least 20 minutes of therapy and no more than 45 minutes of therapy 
Medium priority when the last available battery is the external DC source and can provide at least 20 
minutes of therapy. 
High priority when the last available battery can provide at least 10 minutes of therapy.   
Why it occurs 
The last battery available is low or nearly depleted. This includes the internal battery. 
Why it occurs 
The proximal pressure line may be improperly connected or it may contain water droplets that affect 
the pressure reading. 
What to do 
• 
Verify the proximal pressure line is connected properly at both ends and the line is clean and 
untangled. 
• 
If necessary, clear the line and reconnect. 
Requirements 
Proximal pressure line attached 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
Monitored parameters and alarms that use the pressure measurement will not function 
Monitored parameters and alarms that use the flow measurement, such as tidal volume, will continue 
to function 
Algorithm 
summary  
During periods of the breath when there is low flow in the inhalation limb, the pressure measured at 
the outlet of the device is compared to the pressure measured by the proximal line.   When the 
difference is greater than 5 cm H
2
0 at low flow for three consecutive breaths, the alarm occurs.