Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Trilogy Evo Universal Accessories 
 
 
 
Circuit Accessories 
81 
 
Patient Circuits 
For instructions on attaching circuits to the ventilator, see the “Device Setup” chapter. 
Circuit Principles 
Each patient circuit has a different compliance and resistance. During Circuit Calibration, the ventilator verifies that the maximum pressure drop 
does not exceed the specifications below. For optimal performance, calibrate the circuit before using it with a patient.  See “Circuit Calibration” in 
the “Device Options” chapter.  
When adding any components to the breathing system, the flow resistance and dead space of the added components such as humidifiers, speaking 
valves, Heat Moisture Exchangers (HMEs) and filters should be carefully considered in relation to the potential for adverse effects on the patient’s 
ventilator management and device alarms. 
If a breathing system filter is exposed to nebulization or humidification, to prevent increased resistance or blockage, the breathing system filter 
requires replacement more frequently. 
Circuit Requirements 
For safe operation, the ventilator requires a patient circuit and filter(s) that meet the following requirements. 
Compliance: up to 4 ml/cmH
2
 
Inspiratory/Expiratory Resistance 
• 
Pediatric/adult circuit (20-22 mm): less than 5 cmH
2
O at 30 l/min 
• 
Pediatric/adult circuit (19 mm): less than 5 cmH
2
O at 15 l/min 
• 
Infant/pediatric circuit (14-16 mm): less than 5 cmH
2
O at 15 l/min 
• 
Infant/pediatric circuit (9-13 mm): less than 5 cmH
2
O at 2.5 l/min 
Leak Compensation 
The passive circuit provides leak compensation for inhaled and exhaled tidal volume measurements.  This includes compensation for intentional 
leak in the patient circuit and leaks that occur at the patient interface, such as cuff leak or mask leak.   
The active flow and dual limb circuits compensate for leak that occurs between the ventilator and the external flow sensor.  Leaks that are 
downstream of the external flow sensor are not compensated for in measurement of inhaled or exhaled tidal volume. 
Leak compensation is not available in the active PAP circuit. 
Circuit Accessories 
HMEs 
Three Heat and Moisture Exchangers (HMEs) are included in this package: adult, pediatric, and infant.  
Follow the instructions provided with the sensors.  
 
Flexible Trach Adapter 
Adult Endo-trach adapter with port for suctioning. Follow the instructions provided with the adapter.