Specifications 
90 
 
Monitored Parameter Accuracy 
Airway pressure  
±(2 cmH
2
O + 4% of actual) 
Tidal volume 
 ± (4 ml + 15% of actual) for volumes ≥ 35 ml 
± 10 ml for volumes < 35 ml 
FiO
2
 
±(2.5% FiO
2
 + 2.5% of actual reading) within a 24-hour 2-point calibration 
period, or a change in altitude 
EtCO
2
 (with Mainstream CO
2
 sensor) 
±(volume fraction of 0.43%  + 8% of gas level) 
SpO
2
 and pulse rate 
See the sensor manufacturer’s operating instruction sheet. 
Operational range of the Patient Circuit and Circuit accessories 
Inspiratory/expiratory resistance 
• 
Pediatric/adult circuit (20-22 mm): less than 5 cmH
2
O at 30 l/min 
• 
Pediatric/adult circuit (19 mm): less than 5 cmH
2
O at 15 l/min 
• 
Infant/pediatric circuit (14-16 mm): less than 5 cmH
2
O at 15 l/min 
• 
Infant/pediatric circuit (9-13 mm): less than 5 cmH
2
O at 2.5 l/min 
Compliance 
Up to 4 ml/cmH
2
Wireless 
Bluetooth
 
Operating frequency range
 
2402-2480 MHz
 
Channel bandwidth 
 
1 MHz/2 MHz
 
Maximum output power
 
12.0 dBm
 
Modulation
 
GFSK, Pi/4 DQPSK, 8DQPSK
 
Near-field communication (NFC) 
Operating frequency 
13.56 MHz 
Receiving section bandwidth  
1.4 MHz 
Maximum output power 
23.0 dBm 
Modulation 
ASK, OOK 
Wi-Fi 
Operating frequency range 
2402-2480 MHz 
Channel bandwidth 
20 MHz/40MHz 
Maximum output power 
17 dBm 
Modulation 
DSSS, OFDM, DBPSK, DQPSK, CCK, 16-QAM 
Security 
Wpa2 
All flows and volumes are expressed in BTPS. 
Measurement Uncertainty for Control and Performance Specifications  
The stated tolerances account for the measurement uncertainty of the test equipment used to verify performance:  
Pressure: ± 0.75% cmH
2
Tidal Volume: ±2 ml 
Oxygen: ±1% FiO
2