Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Technical Specifications 
 
 
 
EMC Information 
92 
 
Pneumatic Diagram 
 
AECM
Filter
Dual 
Limb / 
PAP
Air Inlet
Low 
Flow 
O2
OBM
O2 Flow 
Sensor
High 
Flow 
O2
O2 Press   
Inlet Flow Path
FiO2 
Sensor
Outlet Flow Path
Machine Flow 
Sensor / 
Element
System PCA
Monitor 
Pressure
Outlet   
Pressure
Blower
 
EMC Information 
This ventilator generates audible and visual alarms to alert you if it is not able to provide ventilation or if external monitoring is 
lost during an EMC disturbance. 
Warnings 
Avoid using this equipment adjacent to or stacked with other equipment because it could result in improper operation. 
Although the other equipment may comply with EMC standard requirements, interference can occur. If such use is necessary, 
observe this equipment and the other equipment to verify that both are operating normally. 
This device is unsuitable for use in a magnetic resonance imaging (MRI) environment. 
The use of accessories, transducers and/or cables other than those specified, with the exception of those sold by the 
manufacturer as replacement parts for internal components, may result in increased emissions or decreased immunity of the 
equipment or system.