BioSensics LLC PM002 Manual Do Utilizador

Página de 17
PAMSys User’s Manual V2.0 
Data Collection
 
 
 
For PAMSys-X, the multi-sensor configuration must first be synchronized before beginning to 
record data. Press the Reset/Power button on the front face to power each of the five sensors, with 
the CHEST sensor being turned on last. The LED will quickly blink orange when the sensor is on. 
Once all the sensors are on, the CHEST sensor briefly shows a red light, then they will attempt to 
sync amongst each other. The sensors are successfully synced when the CHEST sensor blinks 
orange independently, and slightly before, the WRIST and LEG sensors, which will blink orange 
simultaneously. This sequence will occur every three seconds. 
 
Recording Data with PAMSys 
To begin recording data, turn the sensor on by pressing the reset/power button. Next, the user should 
stand up and place the sensor on the body using the chest strap or PAMSys shirt as directed above. 
If using the PAMSys-X system please synchronize the sensors before beginning to record data. The 
user must stand in place for at least 10 seconds. After this period of time, the sensor is now ready to 
collect and record physical activity data. The time when a new measurement is started and when 
the subject is standing still and upright should be noted. These times will be used in 
PAMWare to properly analyze the data (see ‘Data Analysis’).