BioSensics LLC PM002 用户手册
PAMSys User’s Manual V2.0
9
Data Collection
For PAMSys-X, the multi-sensor configuration must first be synchronized before beginning to
record data. Press the Reset/Power button on the front face to power each of the five sensors, with
the CHEST sensor being turned on last. The LED will quickly blink orange when the sensor is on.
Once all the sensors are on, the CHEST sensor briefly shows a red light, then they will attempt to
sync amongst each other. The sensors are successfully synced when the CHEST sensor blinks
orange independently, and slightly before, the WRIST and LEG sensors, which will blink orange
simultaneously. This sequence will occur every three seconds.
Recording Data with PAMSys
To begin recording data, turn the sensor on by pressing the reset/power button. Next, the user should
stand up and place the sensor on the body using the chest strap or PAMSys shirt as directed above.
If using the PAMSys-X system please synchronize the sensors before beginning to record data. The
user must stand in place for at least 10 seconds. After this period of time, the sensor is now ready to
collect and record physical activity data. The time when a new measurement is started and when
the subject is standing still and upright should be noted. These times will be used in
PAMWare to properly analyze the data (see ‘Data Analysis’).