Innokas Yhtyma Oy VC150 Manual Do Utilizador

Página de 332
KO00065K
VC150 Vital Signs Monitor
B-7
Maintenance: Cleaning
Cleaning the Welch Allyn probe and probe well
Units that are in service on a regular basis should have the following preventive 
maintenance performed every six months:
1. Visually inspect the thermometer probe for any physical damage that might 
cause future product failure.
2. Clean the probe and the probe well according to instructions below. Use an 
alcohol-based cleaning agent on the Welch Allyn metal probe or plastic 
section of the probe only.
Cleaning the Welch Allyn probe
As needed, clean the probe with a 70% isopropyl alcohol solution or a 10% 
chlorine bleach solution or a nonstaining disinfectant such as CaviCide
®
 or its 
equivalent.
WARNING
Do not immerse or soak the probe in any type of fluid. Do not use steam, 
heat or gas sterilization on the probe. Do not autoclave the probe.
CAUTION
Never immerse any monitor accessories.
Cleaning the removable probe well
1. Remove the probe well from the unit.
2. Unplug the latching probe connector to prevent the monitor 
from consuming battery power while you are cleaning the 
probe well.
3. Clean the inner surface of the probe well by swabbing the 
surface with a cloth dampened with a mild detergent 
solution or a 70% isopropyl alcohol, or a 10% chlorine bleach solution, or a 
nonstaining disinfectant such as CaviCide
®
 or its equivalent.
4. Clean the probe well’s outer surface by swabbing or wiping the surface with 
one of the solutions mentioned above. Immerse the probe well in mild 
detergent solution as necessary for cleaning.
5. Thoroughly dry all surfaces before re-assembling the instrument.
6. Reconnect the latching probe connector to the monitor. Ensure that the 
connector snaps into place.
7. Reinstall the probe well in the monitor and snap the probe well into place.
8. Insert the probe into the probe well.