The undesired signals can be filtered out by setting the coupling mode. For example,
the signal under test is a square waveform with DC offset.
When the coupling mode is “DC”: the DC and AC components of the signal under
test can both pass the channel.
When the coupling mode is “AC”: the DC components of the signal under test
are blocked.
When the coupling mode is “GND”: the DC and AC components of the signal
under test are both blocked.
Press CH1 Coupling and use
to select the desired coupling mode (the
default is DC). The current coupling mode is displayed in the channel label at the
bottom of the screen. You can also press Coupling continuously to switch the
coupling mode.
Bandwidth Limit
Setting the bandwidth limit can reduce the display noise. For example, the signal
under test is a pulse with high frequency oscillation.
When bandwidth limit is disabled, the high frequency components of the signal
under test is a pulse with high frequency oscillation.
When bandwidth limit is disabled, the high frequency components of the signal
under test can pass the channel.
Enable bandwidth limit and limit the bandwidth to 20 MHz or 100 MHz (only
applicable to 200 MHz and 300 MHz oscilloscopes), the high frequency
components that exceed 20 MHz or 100 MHz are attenuated.
components that exceed 20 MHz or 100 MHz are attenuated.
Note: for DS2102 and DS2072, the bandwidth limit can only be set to 20 MHz.
Press CH1 BW Limit and use
Press CH1 BW Limit and use
to enable or disable bandwidth limit (the
default is OFF). When bandwidth limit (20 MHz or 100 MHz) is enabled, the character
“B” will be displayed in the channel label at the bottom of the screen. You can also
press BW Limit continuously to switch between on and off of the bandwidth limit.