DS2000A User’s Guide
 
 
2-3  
 
Channel Coupling   
 
The undesired signals can be filtered out by setting the coupling mode. For example, 
the signal under test is a square waveform with DC offset. 
  When the coupling mode is “DC”: the DC and AC components of the signal under 
test can both pass the channel. 
  When the coupling mode is “AC”: the DC components of the signal under test 
are blocked. 
  When the coupling mode is “GND”: the DC and AC components of the signal 
under test are both blocked. 
 
Press CH1  Coupling and use 
  to select the desired coupling mode (the 
default is DC). The current coupling mode is displayed in the channel label at the 
bottom of the screen. You can also press Coupling continuously to switch the 
coupling mode.   
 
 
Bandwidth Limit   
 
Setting the bandwidth limit can reduce the display noise. For example, the signal 
under test is a pulse with high frequency oscillation. 
  When bandwidth limit is disabled, the high frequency components of the signal 
under test can pass the channel. 
  Enable bandwidth limit and limit the bandwidth to 20 MHz or 100 MHz (only 
applicable to 200 MHz and 300 MHz oscilloscopes), the high frequency 
components that exceed 20 MHz or 100 MHz are attenuated. 
Note: for DS2102 and DS2072, the bandwidth limit can only be set to 20 MHz. 
 
Press CH1  BW Limit and use 
  to enable or disable bandwidth limit (the 
default is OFF). When bandwidth limit (20 MHz or 100 MHz) is enabled, the character 
B” will be displayed in the channel label at the bottom of the screen. You can also 
press BW Limit continuously to switch between on and off of the bandwidth limit.