Lecroy WSXS CANBUS TD WSXS-CANBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® Oscilloscopes. WSXS-CANBUS TD Manual Do Utilizador

Códigos do produto
WSXS-CANBUS TD
Página de 69
Operator's Manual 
 
CANbus-TD-TDM-OM-E RevB 
67 
 
Data Value -- The Data Value is set in Hexadecimal format.  If 
you wish, you can precede the ID value with a “0x”, but this is 
not necessary.  Make sure to enter a Data Value that matches 
the DLC Value.  If you have set the Data Condition to INRANGE 
or OUTRANGE, then you will enter two Data Values. 
If you wish to set the Data Value in a Binary format, reference 
the separate section on how to do this.  
CAN data can be formatted in several different formats – MSB 
(Most Significant Byte) or LSB (Least Significant Byte.  In all 
cases, bit data is lsb (least significant bit, bitwise to the left, or 
the 7
th
 bit in the byte appearing at the far left of the byte, and 
the 0
th
 bit at the far right of the byte).   
The CAN controller used in the Trigger Module formats data in 
MSB and lsb (bitwise to the left) format.   
Note: The decoding algorithm always decodes the CAN signal 
(in hexadecimal nibbles and bytes) using an LSB and lsb 
condition.   
 
Start Bit and # Bits -- The CANbus TD trigger allows you to 
trigger on up to 24 contiguous data bits (3 data bytes).  This 
maximum 24-bit string can start at any location in the CAN 
message data field --- it is not limited to the start of a full byte 
or a nibble.   
The Start Bit can be any value from 0 to 63.  If you enter a value 
less than 0, it will default to 0.  If you enter a value more than 
63, it will default to 63.  The Start Bit value is always in LSB 
format (i.e., the bit number as shown on the decoded 
waveform, with bit 0 being at the far left and bit 63 being at the 
far right of the data string).  Remember that the 1
st
 data byte is 
bits 0-7, the 2
nd
 data byte is bits 8-15, etc.  Also, make sure that 
your Start Bit value makes sense in relation to the DLC Value.  
For instance, a Start Bit value of 32 with a DLC Value of 4 is not 
going to result in a successful trigger.   
Reference the Trigger Setup Examples section for more 
information on setting up a trigger where the start bit is not the