Lecroy WSXS-LINBUS TD WSXS-LINBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-LINBUS TD Manual Do Utilizador

Códigos do produto
WSXS-LINBUS TD
Página de 204
Operator's Manual 
919586 RevA 
81 
 
Frame - When selected, the Setup Format, Frame ID Setup, Cycle Count, and Frame Qualifiers fields are 
shown and enabled. 
Note: Some Frame ID Setup and Cycle Count fields are enabled based on selected condition values 
indicated in respective detail sections). 
 
Symbol - When selected, the Setup Format, Frame ID Setup, Cycle Count, and Frame Qualifiers fields 
aren't shown (and are therefore disabled). Instead, the Trigger On and System Parameters fields (step 4b 
and 5b, as follows) are shown and enabled. 
 
Errors - When selected, the Setup Format, Frame ID Setup, Cycle Count, and Frame Qualifiers fields aren't 
shown (and are therefore disabled). Instead, the Protocol Errors, CRC Errors, Select all Errors, and 
Deselect all Errors fields (step 4c and 5c, as follows) are shown and enabled. 
Note: When the Payload CRC checkbox on the CRC Errors section is selected, the Payload Channel field 
is enabled. Select a channel as desired. 
TSS
 
(S
TART
)
 OR 
F
RAME 
T
RIGGER 
S
ETUP 
D
ETAIL
 
4a. Setup Format - TSS (Start) or Frame Trigger Setup Only 
When the FlexRaybus trigger selection is TSS (Start) or Frame, select either Binary or Hexadecimal (Hex) 
setup mode. The format propagates through the entire FlexRaybus trigger setup. 
Note: Completely different fields are shown (instead of Setup Format, Frame ID Setup, Cycle Count, and 
Frame Qualifiers sections) when Symbol and Error FlexRaybus triggers are used. 
A user can select Binary mode, and set up the address in binary format, then reselect Hex mode and set 
up the data in hexadecimal format. Toggling back and forth between the modes does not result in lost 
information (binary is used internally as the core format for all triggering and decoding operations), 
though use of don’t care bits in a binary setup results in the display of an X (for a full nibble don’t care) or 
a $ (for a partial nibble don’t care). 
F
RAME 
T
RIGGER 
S
ETUP 
D
ETAIL
 
5. Frame ID Setup - Frame Trigger Setup Only 
When Frame is selected as the FlexRaybus trigger, use the Frame ID Setup fields as follows: 
Note: The Frame ID setup fields are shown, but disabled when TSS (Start) is selected as the FlexRaybus 
trigger. Completely different fields are shown in these sections when Symbol and Error FlexRaybus 
triggers are used. 
 
Condition - Select from EqualNot EqualLess thanLess than or Equal toGreather thanGreater than 
or Equal to
In Range, or Out Range conditions available. The default setting is Equal. 
 
Value - Use this field's keypad to enter the desired Frame ID. 
 
To - When the condition is set to In Range of or Out of Range, select a To value specifying the full ID 
range for the trigger. 
6. Cycle Count - Frame Trigger Setup Only 
Cycle Count combines with Frame ID enabling powerful FlexRay triggering. The Cycle Count is a decimal 
value between 0 and 63 correlating to the FlexRay Cycle Count numbering system. The default Value is 
Cycle Count 0 (Value). 
When Frame is selected as the FlexRaybus trigger, use the Cycle Count fields as follows: