Press Esc to fail the Asset and save or abort the results, 
OR 
Press “R” to re-run the continuity test. 
 
  
 
 
3)  Enter the cable Length in metres and the conductor Cross-Section (mm
2
) OR a Resistance can be 
entered if this is known. 
 
4)  Press TEST to complete the test. The PAT400 will now proceed to the next test if the resistance was 
a PASS or to a fail screen if the resistance was still outside the modified limit. 
 
 
See also Section A.6 for cable length limitations and RCD protection on long leads. 
 
 
3.12  WARNINGS  
 
3.12.1  Power-On tests 
 
Running a leakage or load test will operate the asset under test. Some assets, such as vacuum cleaners, 
disc cutters etc can draw high initial load currents (Inrush current). The PAT400 will under these 
circumstances report the asset has a short circuit. 
 
If this warning is displayed, it is recommended the following actions are taken: 
 
1)  Where possible check the asset does not have a short circuit (the fuse would normally be blown in 
such circumstances, prior to it being replaced). 
 
2)  Proceed ONLY if you believe the asset is not a true short circuit, but has a high inrush current. 
 
3.12.2  Reversed supply  
 
UK instruments: 
 
 
The PAT400 will detect a reversed polarity (Phase – Neutral) supply and abort the initialisation. 
 
 
The display will show the warning: 
 
 
 
 
 
ERROR 82: 
Mains polarity reversed 
 
 
 
 
European instruments: 
The instrument will pass initialisation and display the Home screen. The supply polarity will be 
shown as “Reversed”. 
 
 
3.12.3  Supply earth fault