The PAT400 will test for a good supply earth during initialisation. A faulty or missing earth will return the 
error: 
 
 
 
Error 73 – Mains supply Earth Fault 
 
3.12.4  Parallel earth paths 
 
 
 
If the PAT400 detects more than one earth path, the warning: 
 
 
 
Unable to verify earth connection, 
 
 
     Parallel paths may exist. 
 
 
        Refer to user guide. 
 
 
The PAT400 has detected more than one earth path during the Earth continuity, Earth bond, Insulation or 
Earth leakage tests. Refer to Appendix A.8 for further details. 
 
3.12.5  Database full 
 
 
Exceeding the storage capacity of the database will create the following warning: 
 
 
 
Error - Database Full. Unable to save data! 
 
 
 
Press OK to continue 
 
The test result will NOT be saved. 
To free up memory for further test results: 
1.  Backup the existing database to a USB storage device. 
 
2.  Delete records from the PAT400. 
 
See Section 4 for further information on PAT400 memory.