 UV exposure device
Time input
1.
Turn you the knob to the left or to the right.
The first display place (from the right) flashes.
2.
Turn the 
knob 
step-by-step to the right.
In the flashing display place, the value increases according to the rotary motion.
Turn again to the left if the set value is too big. You can also turn to the right until the
desired value is displayed again.
3.
Press onto the knob to take over the set value (in no case, turn the knob
simultaneously).
After this, the next display digit flashes.
In the same way, you can enter the other two values.
All three display digits flash again after the third value has been entered, i.e. the device is
ready for operation.
You can immediately initiate a new setting cycle if the preset time does not correspond to
your desired value. For this, turn the knob again.
Starting of the exposure
Switch the UV-lamps on only when the lid is closed!
Press the rotary knob.
The timer and the fluorescent tubes are turned on. The flashing of the display changes to a
continuous lighting and the set count-down is 
started.
You can interrupt the count-down at any time by pressing the rotary knob.
The display flashes again and the fluorescent tubes are switched off,
The remaining time remains in the display. The count-down continues if you press the rotary
knob now again.
At the end of the count-down, the fluorescent tubes are switched-off. Now, the value set
before the start of the timer reappears flashing in the display.
3.4
Replace of a fluorescent lamp
To replace a flourescent lamp in the top, you only have to unscrew the two Alu-ledges and
to remove the glass pane. Turn the lamp 90” in the socket and take it out.
To replace a flourescent lamp in the bottom, you have to push down on the glass pane
carefully till it bottoms out, and pull it forward till the back of the pane clears the frame.
Then pull the glass pane out at the back side.
Please be careful not to damage the plastic tube leading from the vacuum pump!
To replace the lamp (turn 90”) and reinstall it in reverse order.
7