Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Ficha De Dados
![Bk Precision](https://files.manualsbrain.com/attachments/016dad641fa59d615baa4a964945c7c21f1472c8/common/fit/150/50/8eb4cde39dcb23dd40e21f69d37ce95d3e2d50161395bf7929226a210d8c/brand_logo.gif)
Códigos do produto
BK-8500
Byte
offset
Meaning
3 is constant resistance (CR)
4-24
Reserved
0x3B Read the list operation (CC/CV/CW/CR)
Byte
offset
Meaning
3
List operation mode:
0 is constant current (CC)
1 is constant voltage (CV)
2 is constant power (CW)
3 is constant resistance (CR)
1 is constant voltage (CV)
2 is constant power (CW)
3 is constant resistance (CR)
4-24
Reserved
0x3C Set how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3
How lists repeat:
0 is ONCE
1 is REPEAT
1 is REPEAT
4-24
Reserved
0x3D Read how lists repeat (ONCE or REPEAT)
Byte
offset
Meaning
3
How lists repeat:
0 is ONCE
1 is REPEAT
1 is REPEAT
4-24
Reserved
0x3E Set the number of list steps
Byte
offset
Meaning
3 to 4
2 byte little-endian integer for number of steps
5-24
Reserved
0x3F Read the number of list steps
Byte
offset
Meaning
3 to 4
2 byte little-endian integer for number of steps
5-24
Reserved
8500 DC Load Series
Version: September 2, 2009
Page 60 of 76