Ohaus ohaus indicator cd-11 Manual Do Utilizador

Página de 120
EN-12
CD-11 Indicator
Press the G/N/T/
Menu
      F   xx is displayed.....  This is the full scale capacity
setting.   xx= value last set.  Available selections are are shown in the Setup
Table in section 2.6.3.
Press the PRINT/
Units 
button until desired capacity value is reached.
Press the G/N/T/
Menu
 button, gd0.001 is displayed.  This is the graduation
size setting.  Available selections are shown in the Setup Table in section
2.6.3.
Press the PRINT/
Units
 button until desired graduation value is reached.
Press the G/N/T/
Menu
 button, CP   5 kg is displayed.  This is the full scale
calibration point setting.  The range is from 5kg/lb to 100% Full scale capacity.
Press the PRINT/
Units 
button until desired calibration value is reached.
Press the G/N/T/
Menu
     button to end this block, SAVE is displayed.
Press the G/N/T/
Menu
 button to save the menu setup setting.  The next menu
rEAD is displayed.  The Indicator is now matched up with the scale base and
the Indicator parameters may now be set and calibrated.
or
Press the PRINT/
Units
 button to return to the SEtuP menu without saving
changes.
2.6.4 Setup Menu (Cont.)
Procedure