Ohaus ohaus indicator cd-11 Manual Do Utilizador

Página de 120
EN-13
CD-11 Indicator
2.6.5 Readout Menu
The Readout menu is used to adapt the Indicator to environmental conditions and set various features that
include: averaging level, stability level, measuring units, parts counting, display hold, auto zero tracking, timer,
retain zero data, backlight and a factory service mode.  Review all of the settings available before proceeding.
Procedure
To select any of the items in the Readout menu, proceed as follows:
NOTE: If entering from the preceeding menu, disregard the first step.
With the Indicator ON, press and hold the     G/N/T/
Menu
 button until MEnU is
displayed.  When the G/N/T/
Menu
 button is released, CAL is displayed, then
press the     PRINT/
Units
  button until rEAd is displayed.
Press the G/N/T/
Menu
     button, rESEtn is displayed.  This allows resetting the
rEAd menu to factory defaults.
rESEtn = no, does not reset settings.
rESEty= yes, will reset the entire readout menu to factory defaults as
follows:
AL Lo, StAb 1, UnOff g, Un On kg, Un On Lb, Un Off oz, Alt Off, AZt 0.5,
AOt Off, rZd Off, bLAuto and EP OFF (service mode).
Press the PRINT/
Units
     button for selections N or Y.
AVERAGING LEVEL
Press the G/N/T/
Menu
     button, AL  LO is displayed.  This is the averaging level
settings.  Selections are:
 'Lo' - Less processing, less stability and faster stabilization time.
 'Hi' - More processing, greater stability and slower stabilization time.
  (This is the default setting)
Averaging level compensates for vibration or excessive air currents on the
scale base.  During operation, the Indicator continually takes weight readings
from the load cell.  Successive readings are then digitally processed to achieve
a stabilized display.  Using this feature specifies how much processing is
needed.
Press the PRINT/
Units
     button until desired averaging level setting is reached.