Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: This procedure will bring the boom down or lift it up if the hand control is not functioning 
properly.
To activate the primary emergency release, insert a pen into the hole labeled Emergency 
Up or Emergency Down on the control box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: All patient lift actuators are equipped with a mechanical emergency release. The 
mechanical release will enable the actuator to retract without power.
NOTE: Use the primary emergency release first before using the secondary emergency release 
procedure. This procedure should only be used if the primary emergency release procedure is not 
functioning or is unreachable.
NOTE: The lift MUST be under a load for the mechanical release to function.
To activate the secondary release, pull up on the EMERGENCY ring and push down on 
the boom at the same time.
EMERGENCY
Emergency
Release 
Hole
Control 
Box
Pull UP on 
EMERGENCY 
Ring.
Push DOWN 
on Boom