SECTION 3—OPERATION
18
Charging the Battery
NOTE: For this procedure, refer to FIGURE 3.5.
NOTE: Invacare recommends the battery be recharged daily to prolong battery life.
NOTE: An audible alarm will sound when battery is low.
1. Lift up on the handle on the back of the battery.
2. Lift the battery up and away from the control box.
CAUTION
Make sure there is an audible click when mounting battery on the battery charger to 
ensure proper mounting. Otherwise, injury or damage may occur.
3. Place the battery on the battery charger. Push the top of the battery against the mounting 
bracket until there is an audible click.
NOTE: The charge LED will illuminate. When charged, the LED will stop illuminating.
NOTE: It will take approximately four hours to charge a battery that requires a full charge.
4. Lift up on the handle on the back of the battery.
5. Lift the battery up and away from the battery charger.
CAUTION
Make sure there is an audible click when mounting battery onto the control box to 
ensure proper mounting. Otherwise, injury or damage may occur.
6. Place the battery on the control box. Push the top of the battery against the mounting 
bracket until there is an audible click.
FIGURE 3.5   Charging the Battery
Battery 
Charger 
(STEP 3)
Control Box 
(STEP 6)
Audible click 
when properly 
installed
(STEPS 3 and 6) 
Handle
(STEPS 1, 2,  4 and 5)
Battery