DIAGNOSTIC MODE
HI 93114 facilitates operations by providing a
diagnostic mode. In this mode, user can set
or verify different parameters necessary to
ensure optimum performance of the meter.
To enter the diagnostic mode, turn the meter
on and momentarily press ALT and CAL to-
gether. The display will show four dashes to-
gether with “d 00”:
Using the  ,   and 
 keys, select the re-
quired diagnostic mode and press the 
 key.
The meter will execute one of the following
user-diagnostic functions:
10
Customizes automatic shutdown
11
Selects User or Factory functions
12
Turns logging mode on or off
21
Calibrates span in turbidimetric mode
31
Calibrates span in colorimetric mode
40
Clears the logged memory
The following  diagnostic modes are reserved
for authorized service technicians:
00
Shows the Blank level in colorimetric mode
01
Shows the Sample level in colorimetric mode
02
Shows the Dark level in colorimetric mode
05
Shows the Ground voltage
06
Shows 5V on-board level
07
Shows battery voltage level
08
Shows 1.23V reference voltage level
09
Shows -5V on-board level
99
Shows software version number
Subsequently, the up-
per display will show
“----”, indicating that the
meter is calibrated and
ready to measure the
concentration of an un-
known sample.
By pressing the CAL key dur-
ing calibration, the calibration
mode can be exited at any time
without changing the previously
stored calibration data.
+