Hanna Instruments Turbidity & Chlorine Measurements HI 93114 Manual Do Utilizador

Página de 19
29
28
of the display to remind the user that every time
a measurement is taken, the value is stored in
the next available lot number.
REVIEWING THE LOGGED BUFFER
To review the memo-
rized values, press to-
gether ALT and  .
The meter will scroll all the data in the buffer
showing the lot number, value, date and time.
e.g. The first recorded reading in the buffer is
lot # 0, 0.35 mg/L of Free Chlorine,
memorized on 23
rd
 August at 3:34 pm;
The second logged data relates to lot
# 1, 1.35 mg/L of the customized param-
eter, logged on 23
rd
 August at 3:55 pm.
First lot
Second lot
CLEARING THE MEMORY
After all the buffer (memory) is taken up, the
LCD will blink “Full”.
LOGGING
HI 93114 allows user to log 25 time/day-tagged
measurements. User can easily turn the log-
ging mode on and off, review the logged
memory, review the current lot number and
clean the buffer (memory). HI 93114 also
reminds the user if its memory is full.
TURNING THE LOG MODE ON OR OFF
Enter the diagnostic
mode by pressing ALT
and CAL together.
Select mode 12 and press the
 key.
The display will show the current (vacant) lot
together with “LOG” if the log mode is on.
Otherwise it will show “----” if the log mode is
off.
By pressing the 
 key, the meter toggles
between the Log on and off positions. If the
log on mode is selected, every time a mea-
surement is taken (READ pressed) the rel-
evant values will be stored in the current
(vacant) lot number.
To quit diagnostic
mode, press ALT and
CAL together again.
The LCD will then show “CAL” and “Stor” for a
few seconds. If the log on mode was selected
“LOG” will appear on the bottom left hand side
+
To quit diagnostic mode,
press the ALT and CAL
keys together again.
+
+
+