GDC-150 Operation Manual 
 
 
16 
 
5209-9000 Rev 0 
 
Figure 4-2.  Calculating the Alarm Set Point 
4.7.2.  Adjusting the Alarm Set Point 
 
Step 
Description 
1. 
Move the jumper to position  P5.  The green outer LED will flash once 
within 2 seconds for confirmation.  The system is now waiting for the user 
to set the desired value. 
2. 
Attach  volt  meter leads to test points TP1 and TP2  (see  Figure  1-4  on 
page 3).  Alternatively, if the optional LED digital display has been fitted, 
read the value indicated on the display. 
3. 
Use  the UP or DOWN push buttons  until  the calculated voltage is 
displayed on the volt meter (or on the optional LED digital display). 
4. 
Move the jumper back to its resting position (P1).  At this time the new 
value is saved and the green LED flashes once for confirmation. 
 
5.  CALIBRATION 
 
5.1.  Frequency Recommendations 
 
Sensor Types / Applications 
Calibration Frequency 
Parking garage detectors 
Once every 12 months 
OHS (Occupational Health and Safety) applications 
Once every 6 months 
  
For the purposes of safety in OHS applications, sensors should be gas tested (bump 
tested) once every month to confirm response. 
 
Required Equipment:  
• 
Volt meter 
• 
Calibration kit 
• 
Calibration gases. 
  
Users can order the calibration kit, calibration accessories and/or gases from any 
Bacharach authorized distributor or they can supply their own gas and equipment as 
long as the gas meets the minimum specifications indicated in this manual.