Chapter 7 Move a monitored patient
Welch Allyn Acuity Central Monitoring System
To arrange a patient move from a wireless Propaq CS monitor
1.
From the main menu, touch SETUP, ACUITY, PATIENT.
2.
Take one of these steps:
•
Select NEW ROOM to move the patient to a new room within
the unit, and then select the new room.
the unit, and then select the new room.
The patient is confirmed with the new room.
•
Select TRANSFER to transfer the patient to a different unit, select the new unit,
and then select the new room.
and then select the new room.
The patient is confirmed and the patient’s map icon appears at the new central
station.
station.
3.
(Optional) Switch a wireless Propaq CS monitor to a
hardwired connection:
hardwired connection:
a.
Connect the Acuity System network cable to the Acuity
System network connector on the bedside monitor.
System network connector on the bedside monitor.
b.
Connect the other end of the cable to the bedside network connector.
The patient is confirmed and the patient’s map icon appears in the new location.