Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
Directions for use
Chapter 7   Move a monitored patient
129
Move the patient to a different monitor
You can move patients who are connected to hardwired or wireless monitors to different 
monitors within the same Acuity System network.
Move a patient from a hardwired monitor to a different monitor
1.
On a Propaq CS monitor: Press NET OFF in the top left corner of 
the screen
On a Propaq Encore monitor: Press NET OFF on the top right 
corner of the casing.
On a Welch Allyn 1500 Patient Monitor: Press the Standby key and 
select Discontinue Monitoring.
2.
Turn off the monitor.
The trend information for this patient is erased from the monitor.
3.
Connect the patient to a new monitor, and connect the monitor to 
the Acuity System.
At the central station, the patient’s yellow icon appears in an unconfirmed state on 
the map in the new location.
4.
Enter or confirm the patient information. 
For instructions, see 
Move a patient from a wireless monitor to a different monitor
To move a patient from a Micropaq monitor to a different monitor
1.
On the Micropaq monitor, press 
 to highlight ACUITY, and then press 
 to select it.
2.
Select END TELE.
The monitor disconnects from the Acuity System.
3.
Remove the battery from the Micropaq monitor.
4.
Connect the patient to a new monitor, and connect the monitor to the Acuity System.
At the central station, the patient’s yellow icon appears in an unconfirmed state on 
the map.
5.
Enter or confirm the patient information. 
For instructions, see