Directions for use
Chapter 9   Status icons
137
Patient/monitor status icons and recommended action
This table illustrates these things about patient/monitor status icons:
Display priority
Location on the central station map
Messages at the top of the Acuity Central Station screen
Messages in message panels (when present)
Recommended action
Table 8.  Icons and messages that represent patient and monitor status
Patient/monitor 
status icon
Icon display 
priority
Icon location on map
Messages in 
hallway message 
panels and at top 
of Acuity Central 
Station screen
Recommended action
Port Icon
Locations representing 
unused ports available 
for monitoring Acuity 
System patients.
None
When you are ready, begin 
monitoring from this 
location.
Monitor Connecting
Locations representing 
monitors establishing 
communication with 
Acuity System.
None
Wait until communication 
is established before you 
open the Virtual Monitor 
or Waveform Window.
Monitor Disconnecting
Locations representing 
monitors disconnecting 
from Acuity System.
None
If you have pressed the 
NET OFF key, unplug 
monitor from Acuity 
System within 15 seconds 
or Acuity System 
reconnects.
Patient Intentionally 
Disconnected
Locations representing 
monitors from which 
patients have been 
intentionally 
disconnected.
This icon also appears 
during Manual NIBP-Only 
monitoring and spot 
SpO2 monitoring, but 
does not affect alarming 
at Acuity.
None
If communication between 
the patient and Acuity 
System is reestablished, 
the CONFIRM PATIENT? 
prompt appears. 
Click YES to reconnect the 
original patient to Acuity 
System.
Patient Monitoring OK
Beneath alerts 
and alarms
Locations representing 
patients/monitors that 
are not in an alarm or 
alert state.
None
Auto NIBP-Only 
Monitoring
Beneath alerts 
and alarms
Locations representing 
monitors that are only 
monitoring NIBP, 
automatically.
None