Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
138
Chapter 9   Status icons
Welch Allyn  Acuity Central Monitoring System
Alarm Tone Suspended 
with No Violation
Beneath alerts 
and alarms
Locations representing 
monitors with audible 
alarm tones suspended 
when no alarm or alert 
violation has occurred.
None
At the patient monitor, 
press the Silence/Reset 
key when ready.
Or
At the central station, 
click Resume in the 
patient’s Virtual Monitor 
when ready.
color reflects current 
status of monitor
Wireless
When the View Wireless 
button is pressed in the 
lower left corner of the 
Acuity Central Station 
screen, this icon appears 
in map locations where 
wireless monitors are 
actively communicating 
with Acuity System.
None
Unconfirmed Patient ID
Low-level alert
Locations representing 
monitors with 
unconfirmed patient IDs.
None
In the patient’s Virtual 
Monitor, click Setup
Patient ID
Confirm or enter ID 
information 
Unconfirmed Room 
Number
Low-level alert
Wireless monitor map 
areas, representing 
patients with unassigned 
or unconfirmed room 
numbers. 
None
In the patient’s Virtual 
Monitor, click Setup
Patient ID
Confirm or enter a room 
number.
Some Alarm Limits Off
Low-level alert
Locations representing 
monitors with alarming 
for some vital signs set 
to off.
None
In the patient’s Virtual 
Monitor, click Setup
Alarms
Set at least one upper and 
lower “limits-pair” per 
active vital sign.
Table 8.  Icons and messages that represent patient and monitor status
Patient/monitor 
status icon
Icon display 
priority
Icon location on map
Messages in 
hallway message 
panels and at top 
of Acuity Central 
Station screen
Recommended action