Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
Directions for use
Chapter 9   Status icons
141
Patient Alarm 
or
Telemetry Dropout 
Take priority 
over alerts. 
Locations representing 
monitors that are audibly 
alarming in response to 
one of these conditions:
Patient vital sign 
limit violations
Patient monitor is 
not communicating 
with the Acuity 
System.
[room #] 
VT
ASYS
VF
APNEA
Limit vital sign
or 
[room #] Dropout
Patient Alarm
Attend to the patient and 
suspend or silence the 
alarm tone. You can turn 
off or reset alarm limits 
also at the central station. 
Then click the Silence/
Reset 
key at the patient 
monitor. or Resume at the 
central station Virtual 
Monitor.
Telemetry Dropout
During this state, 
Arrhythmia and ST alarms 
are not indicated.
1. Click one of these 
buttons:
SUSPEND to 
suspend the alarm 
tone for 90 seconds
RESET to stop the 
alert and put the 
monitor in the blue, 
“patient intentionally 
disconnected” state, 
2. Bring the wireless 
monitor within range 
of a wireless access 
point. 
3. If the problem 
persists, contact 
qualified service 
personnel.
No Continuous Heart/
Pulse Rate Source
Take priority 
over alerts. 
Locations representing 
monitors that no longer 
receive continuous heart/
pulse rate signals. If 
monitoring began 
without an HR/PR 
source, the icon remains 
blue.
[room #] 
Source HR/PR
Reconnect the HR/PR 
source at the patient 
monitor. 
Silence or suspend the 
audible alarm tone. 
Table 8.  Icons and messages that represent patient and monitor status
Patient/monitor 
status icon
Icon display 
priority
Icon location on map
Messages in 
hallway message 
panels and at top 
of Acuity Central 
Station screen
Recommended action