Chapter 9   Status icons
Welch Allyn  Acuity Central Monitoring System
Alarm Tone Suspended
Only lethal 
arrhythmia 
events interrupt 
the suspension 
period. All 
other newly 
occurring 
alarms are 
indicated only 
visually, and 
the icon image 
remains the 
same. Newly 
occurring alerts 
are not 
indicated.
Locations representing 
monitors for which 
audible patient alarm 
tones have been 
suspended and the alarm 
state is still active.
[room #]
VT
ASYS
VF
APNEA
Limit vital sign
Attend to the problem, 
and then click the 
Silence/Reset key at the 
patient monitor. or 
Resume at the central 
station. 
Table 8.  Icons and messages that represent patient and monitor status
Patient/monitor 
status icon
Icon display 
priority
Icon location on map
Messages in 
hallway message 
panels and at top 
of Acuity Central 
Station screen
Recommended action