Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
Directions for use
Chapter 11   Specifications
175
Acuity System alarm and alert indications
Table 25.  General Acuity System alarm and alert indications
Feature
Specification
Numeric update time after Acuity 
System receives alarm indication from 
a patient’s monitor 
Within 1 s 
Acuity Central Station tone volume 
Installation choice (hourly settings) 
Audible alarm or alert tone delay time 
to patient’s monitor 
10 s (configurable up to four minutes, 15 seconds) 
Audible alarm or alert tone 
suspension time and activation 
90 s, activated individually for each patient 
Note, frequency range and pause 
length for different types of tones
Monitor equipment alerts: 1 A
4
 @ 440.0 Hz, pause 2.45 s 
Low-level alarms: 3 E
5
 @ 660.0 Hz (1 s apart), pause 1.25 s 
Medium-level alarms: 1 A
5
 @ 880.0 Hz coupled with 1 C#
5
 @ 554.0 Hz, 
pause 0.5 s 
High-level alarms: 3 A
@ 880.0 Hz, pause 5/6 s 
Lethal alarms: 4 A
5
 @880.0 Hz with harmonics, pause 0.8 s
Latched alarms
Visual latching: When feature enabled, the most recent, highest-priority 
arrhythmia alarm latches in the Waveform Window, even after alarm condition 
ceases. Can be configured to apply to lethal/high-level or lethal/high-level/
medium-level arrhythmia alarms. 
Audible latching: Lethal arrhythmia events can be configured to also latch 
audibly. Audio tone latching default is off.
Table 26.  Alarm and alert indications at the Acuity Central Station 
Alarms
(red)
Monitor equipment alerts
(yellow)
System-wide 
equipment 
alerts 
(yellow)
Patient 
information 
alerts
(yellow)
Duration
Indications remain until 
condition is cleared or 
alarm is turned off. 
If latched alarm 
capability is enabled, 
lethal and high-level 
arrhythmia alarms (and 
medium-level arrhythmia 
alarms if so configured) 
persist in Waveform 
Windows until cleared 
at Central.
Indications remain until the 
condition is cleared or the 
alert is acknowledged 
(disabled) at the patient 
monitor (some alerts do not 
allow acknowledgement). 
Certain wireless alerts can 
be RESET at the Acuity 
Central Station. 
Indications remain until condition 
is cleared except Disk Failure 
equipment alert is displayed for 
two minutes, four times an hour.