Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
Directions for use
Chapter 11   Specifications
177
Table 27.  Alarm and alert indications at the Acuity System Virtual Monitor
Alarms
(red)
Monitor equipment alerts
(yellow)
System-wide 
equipment 
alerts
(yellow)
Patient 
information 
alerts
(yellow)
Display Priority Virtual Monitors in alarm 
“bump” those not in alarm, 
when necessary. 
Virtual Monitors in alert state 
are not automatically 
displayed.
No displays in 
Virtual Monitor. 
Virtual Monitors in 
alert state are not 
automatically 
displayed.
Indications
Violating vital sign’s numeric 
is highlighted. +++ means 
measured values fall above 
the monitor’s range, 
--- means they fall below, and 
??? means they could not be 
measured.
Specific message appears in 
border.
Status icon. 
Equipment Alert status 
window appears in Virtual 
Monitor.
For most, a specific message 
appears in border.
Status icon. 
No indications in 
Virtual Monitor. 
Specific message 
in border.
Status icon. 
Message 
Priority
Displays highest level 
message for each patient.
If more than one monitor 
equipment alert occurs for 
one monitor at once, a 
“Multiple” message appears 
in border.
No messages in 
Virtual Monitor. 
Alarms and 
monitor 
equipment alerts 
take priority over 