Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador
Directions for use
Chapter 11 Specifications
181
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The Acuity System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Acuity System should assure that it is used in such an environment.
the Acuity System should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment—
Guidance
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV air
±6 kV contact
±8 kV air
±8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
supply lines
±1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
typical commercial or hospital environment.
Surge
IEC 61000-4-5
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital environment.
typical commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
<5% U
t
(>95% dip in U
t
)
for 5 sec
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
<5% U
t
(>95% dip in U
t
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Acuity System requires
continued operation during power mains
interruption, it is recommended that the
monitor be powered from an uninterruptible
power supply or a battery.
typical commercial or hospital environment.
If the user of the Acuity System requires
continued operation during power mains
interruption, it is recommended that the
monitor be powered from an uninterruptible
power supply or a battery.
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
Note
U
t
is the AC mains voltage prior to application of the test level.