Chapter 11   Specifications
183
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the 
Acuity System
The Acuity System is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the Acuity System can help prevent electromagnetic interference by maintaining 
a minimum distance between portable and mobile RF communications equipment (transmitters) and the Acuity System 
as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output 
Power of Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = 1.2
 
 
80 MHz to 800 MHz
d = 1.2
 
 
800 MHz to 2.5 GHz
d = 2.3
 
 
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters 
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output 
power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2
These guidelines might not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.
P
P
P