Chapter 11   Specifications
Welch Allyn  Acuity Central Monitoring System
Setup windows
Heart rate (HR)
For patients being monitored for arrhythmias, 
the HR value is derived from the arrhythmia 
library instead of from the patient monitor.
Table 33.  Waveform Window
Features
Factory default
Waveform Window auto 
open 
In waveforms views: 
1. Upon monitor connection to Acuity System if empty window 
2. Upon alarm if open space is available OR replaces a non-alarming Waveform Window 
Number of waveforms 
displayed simultaneously 
Up to two 
Display priority of 
waveforms 
1. ECG
3. P2
5. RESP
7. NIBP
2. P1
4. CO
6. SpO
2
Number of vital sign 
numerics displayed 
Up to four. If four or more monitored at once, the third and fourth alternate in the same 
zone; others appear if they alarm 
Display priority of numerics  1. HR/PR
4. P1
7. RR
2. SpO
5. P2 
8. NIBP
3. Alarming numerics
6. CO
2
 and BR
9. Temperature
Heart rate (HR)
For patients being monitored for arrhythmias, the HR value is derived from the arrhythmia 
library instead of from the patient monitor.
Table 32.  Virtual Monitor 
Features
Customizable default 
Factory default
Table 34.  Setup window
Alarms 
Features
Customizable default Factory 
default
Display priority of vital signs 
Up to three vital signs, in this order of priority: 
1. Vital signs in alarm 
2. Active vital signs (in the order listed in window) 
Alarm limits
Presets values
Presets values
Micropaq monitor: 
HR: 50 to 120 bpm
SpO
2
: 90% to 100%
For all monitors except the Micropaq monitor: Presets 
are the factory default alarm limits of the patient 
monitor, based on patient mode. For further information, 
see monitor directions for use.
Stat Set values
Acuity Systems execute the Propaq Encore monitor 
STAT SET function. For further information, see Propaq 
Encore directions for use
.
Auto Set values
Acuity Systems execute the Propaq Encore monitor 
STAT SET function. For further information, see Propaq 
Encore directions for use
.