Directions for use
Chapter 11   Specifications
185
Virtual Monitor and Waveform Window
Table 32.  Virtual Monitor 
Features
Customizable default 
Factory default
Virtual Monitor auto open 
In map view and full screen review view: 
Upon patient’s monitor connection to Acuity 
System if empty window 
Upon alarm if open space is available OR 
replaces a non-alarming Virtual Monitor 
Patient mode 
(listed in patient monitor 
Setup status window) 
Micropaq monitor: adult
Propaq LT, CS and Encore monitors: Adult, 
pediatric or neonate mode, depending upon 
setting at patient monitor 
Display settings
Acuity System utilizes monitor factory default 
settings. For further information, see the monitor 
directions for use. 
Lead selections
Micropaq and Propaq LT monitors:
5-wire cable:
ECG1 LEAD is Lead II 
ECG2 LEAD is Lead V
RESP LEAD is Lead III
3-wire cable:
ECG1 LEAD is Lead II 
Propaq CS and Propaq Encore monitors:
5-wire cable:
ECG1 LEAD is Lead II 
ECG2 LEAD is Lead V
3-wire cable:
ECG1 LEAD is Lead II 
Welch Allyn 1500 Patient Monitor:
10-wire cable: 
ECG1 LEAD is Lead II
ECG2 LEAD is Lead V
5-wire cable:
ECG1 LEAD is Lead II
ECG2 LEAD is Lead V
3-wire cable:
ECG1 LEAD is Lead II
Number of waveforms 
displayed simultaneously 
Up to three 
Display priority of 
waveforms 
All waveforms on (ECG always displayed if 
monitored): 
1. ECG 1
3. ECG 3
 5. CO
2
 7. 
SpO
2
2. ECG 2
4.P1 and P2  6. RESP
8. NIBP
Number of vital sign 
numerics displayed 
simultaneously 
Eight (if six or more are monitored at once, 
Temperature and NIBP values alternate in the 
same space) 
If CO
2
 active and turned on at the patient 
monitor, BR numerics are shown; second choice 
is RR numerics, when impedance pneumography 
is active