Adjust monitoring settings and 
patient information 
Set up typical monitoring at the Virtual Monitor
This table presents typical monitoring functions that you can set up at the Acuity Central 
Station using the Virtual Monitor.
Table 2.  Typical monitoring functions at the Virtual Monitor
NIBP and SpO
2
 spot checks
Task
Clicks at Virtual Monitor
Notes
Start an NIBP
measurement (except for
a Micropaq monitor).
1. NIBP,
2. START
Can be started also at the patient monitor.
(Propaq LT monitor) Start 
an SpO
spot check
measurement. 
1. SpO
2
,
2. START
SpO
2
 Spot Mode must be turned on at the monitor.
Patient mode and alarms
(Micropaq monitor) 
Change the patient mode.
1. MAIN MENU 
2. MORE 
3. MORE 
4. NEXT 
5. CHANGE
You can change a Micropaq monitor’s patient mode only at 
the Acuity Central Station. For all other patient monitors, you 
can change patient mode only at the patient monitor.
WARNING Always check patient mode in the 
border of the Waveform Window or Virtual Monitor. 
Incorrect patient mode can result in unsuitable 
alarm settings, inaccurate heart rate readings and 
inappropriate arrhythmia and ST-segment analyses. 
Temporarily suppress the 
potential for audible 
alarm tones. 
1. SUSPEND
2. When ready to return to 
normal monitoring:
RESUME
Although any alerts for this patient are displayed, SUSPEND 
suppresses alert tones for 90 seconds.