PN Offset
Pwr Gain
PN Offset
Use this field to enter the PN sequence offset index for your base 
station. Each whole offset is equal to 64 chips (= 52.08 
µ
s). Fractional 
values are rounded off, and can be entered, in increments of 0.015625 
(1 chip). Once entered, this value is shared by all CDMA screens.
This value is used by the analyzer when the 
Even Sec In
 field on the 
CDMA GENERATOR screen is set to 
Enable
NOTE
When the 
Even Sec In
 field is set to 
Not
, changes to the 
PN Offset
 field have no effect on the analyzer.
Screens on which this field is present
CDMA ANALYZER
CDMA GENERATOR
CODE DOM (Code Domain Analyzer): auxiliary controls menu
Pwr Gain
This field is used to set the level into the CDMA analyzer when making 
average power measurements. Levels are set in 6 dB increments. The 
IF signal level is displayed as ADC FS when average power is 
measured, and should be between 
1.0 and 
5 dB for best measurement 
results. 
Auto
 automatically sets the gain for an ADC FS value in the range 
of 
1 to 
5 dB. This is the recommended mode of operation.
Hold
 lets you override the automatic setting to manually set the 
gain.
This field is only displayed when an average power measurement is 
selected. 
Screens on which this field is present
CDMA ANALYZER
Main Menu