HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 4
107
CDMA Measurements - Screens and Control Fields
Pwr Intvl
Pwr Zero
Pwr Intvl
This field determines the length of the data block when measuring 
average power. The data block is a time record over which correlated 
measurements are computed.
Range: 0.25 to 5.00 ms
Screens on which this field is present
CDMA ANALYZER
Pwr Scale
The upper part of this field is used to select the reference level for power 
measurements (relative to the total power of the entire CDMA 
channel). The reference is the top line of the screen and defaults to 
0 dB.
The lower part of this field is used to select the display resolution for 
the power measurement; 1, 2 or 5 dB per vertical division.
The power scale fields are displayed in the 
Marker
 menu when the 
Main
 controls 
Measurement
 field is set to 
Power
 or 
Fast Pwr
.
Screens on which this field is present
COD DOM (Code Domain Analyzer): marker controls menu
Pwr Zero
Selecting this field calibrates (zeroes) the average power meter. This 
operation should be performed immediately before making an average 
power measurement. This field is only displayed if the average power 
measurement is selected. 
Operating Considerations
The 
Auto Zero
 field causes average power to be periodically calibrated 
when set to 
Auto
. If the 
Auto Zero
 field is set to 
Manual
, you must use 
the 
Pwr Zero
 field to zero average power. (See the 
 for more information.)
Screens on which this field is present
CDMA ANALYZER
Main Menu