HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
112
Chapter 4 
CDMA Measurements - Screens and Control Fields
Trig Event
Tune Freq
Trig Event
This field determines when the trigger event to the CDMA analyzer 
occurs. When the selected trigger in this field is qualified by the 
selection in the 
Qual Event
 field, the CDMA analyzer is triggered. 
Choices
27 ms
 is an internally generated clock.
20 ms
 is an internally generated clock.
80 ms
 is an internally generated clock.
2 s
 is an internally generated clock.
Delay
 controls a trigger delay timer, beginning from the positive 
edge of the selected qualifier. A delay of 20 
µ
s to 10,000,000 
µ
s is 
valid. Initially, the time delay will be 100 
µ
s. 
Immed
 triggers immediately after receiving the 
Qual Event
 signal.
Default selection: 
Immed
Screens on which this field is present
CDMA ANALYZER
COD DOM (Code Domain Analyzer): trigger controls menu
Tune Freq
Use this field to enter the center frequency of the CDMA channel you 
are analyzing. This field is displayed only when the 
RF Display
 field 
on the INSTRUMENT CONFIGURE screen is set to 
Freq
.
NOTE
Channel tuning
You can configure the Test Set to allow direct entry of 
channel numbers (instead of entering the frequency) by 
using the 
RF Display
 field on the INSTRUMENT 
CONFIGURE screen. See the 
 for more information.
Screens on which this field is present
CDMA ANALYZER
COD DOM (Code Domain Analyzer): main controls menu
Main Menu