Chapter 4 
CDMA Measurements - Screens and Control Fields
RF In/Ant
Threshold
RF In/Ant
This field selects the Test Set’s RF input. 
CAUTION
The maximum safe level into the ANT IN connector is 
60 mW. Exceeding this level could cause permanent 
damage to the Test Set.
Screens on which this field is present
CODE DOM (Code Domain Analyzer): main controls menu
Threshold
Enter the measurement display threshold for making code domain 
timing and phase measurements. Walsh channels that have power 
levels below the threshold value are not displayed. 
Operating Considerations
Querying these measurements for Walsh channels that are below the 
threshold level using HP-IB will return a default value of 9e99.
Screens on which this field is present
COD DOM (Code Domain Analyzer): auxiliary controls menu